SAGE Open Medical Case Reports (Dec 2017)

A practical method of I-131 thyroid cancer therapy dose optimization using estimated effective renal clearance

  • Brandon A Howard,
  • Olga G James,
  • Jennifer M Perkins,
  • Robert A Pagnanelli,
  • Salvador Borges-Neto,
  • Robert E Reiman

DOI
https://doi.org/10.1177/2050313X17745203
Journal volume & issue
Vol. 5

Abstract

Read online

In thyroid cancer patients with renal impairment or other complicating factors, it is important to maximize I-131 therapy efficacy while minimizing bone marrow and lung damage. We developed a web-based calculator based on a modified Benua and Leeper method to calculate the maximum I-131 dose to reduce the risk of these toxicities, based on the effective renal clearance of I-123 as measured from two whole-body I-123 scans, performed at 0 and 24 h post-administration.