PLoS ONE (Jan 2012)

Fluorescence cell imaging and manipulation using conventional halogen lamp microscopy.

  • Kazuo Yamagata,
  • Daisaku Iwamoto,
  • Yukari Terashita,
  • Chong Li,
  • Sayaka Wakayama,
  • Yoko Hayashi-Takanaka,
  • Hiroshi Kimura,
  • Kazuhiro Saeki,
  • Teruhiko Wakayama

DOI
https://doi.org/10.1371/journal.pone.0031638
Journal volume & issue
Vol. 7, no. 2
p. e31638

Abstract

Read online

Technologies for vitally labeling cells with fluorescent dyes have advanced remarkably. However, to excite fluorescent dyes currently requires powerful illumination, which can cause phototoxic damage to the cells and increases the cost of microscopy. We have developed a filter system to excite fluorescent dyes using a conventional transmission microscope equipped with a halogen lamp. This method allows us to observe previously invisible cell organelles, such as the metaphase spindle of oocytes, without causing phototoxicity. Cells remain healthy even after intensive manipulation under fluorescence observation, such as during bovine, porcine and mouse somatic cell cloning using nuclear transfer. This method does not require expensive epifluorescence equipment and so could help to reduce the science gap between developed and developing countries.