East European Journal of Physics (Mar 2022)
Using a Scanner to Measure Absorbed Doses with Radiochromic Film Dosimeters
Abstract
The article provides a sequence of steps for using RISØ calorimeters for calibration and subsequent use of B3 radiochromic film dosimeters (GEX corporation) and a scanner for measuring absorbed doses. Calibration was carried out with the help of electron beam accelerator in the range of absorbed doses of 3 – 40 kGy (measurement range of RISØ calorimeters). In the course of the work, the following was carried out: – calibration of B3 radiochromic dosimetry films using RISØ calorimeters; – plotting a calibration curve for B3 radiochromic dosimetric films; – calculation of approximation functions; – development of a technique for using a flatbed scanner to measure absorbed doses; – estimation of the measurement uncertainties of absorbed doses. Accelerator operation parameters: scanning frequency of the accelerated electron beam – 5 Hz, pulse frequency – 120 Hz, electron energy – 5 MeV, electron beam current – 60 μA. The measurement error of the absorbed dose is 5.8 %.
Keywords