Sensors (Apr 2005)

On Calibration of pH Meters

  • Da-Ming Zhu,
  • K. L. Cheng

DOI
https://doi.org/10.3390/s5040209
Journal volume & issue
Vol. 5, no. 4
pp. 209 – 219

Abstract

Read online

The calibration of pH meters including the pH glass electrode, ISE electrodes,buffers, and the general background for calibration are reviewed. Understanding of basicconcepts of pH, pOH, and electrode mechanism is emphasized. New concepts of pH, pOH,as well as critical examination of activity, and activity coefficients are given. Theemergence of new solid state pH electrodes and replacement of the salt bridge with aconducting wire have opened up a new horizon for pH measurements. A pH buffer solutionwith a conducting wire may be used as a stable reference electrode. The misleadingunlimited linear Nernstian slope should be discarded. Calibration curves with 3 nonlinearportions for the entire 0—14 pH range due to the isoelectric point change effect areexplained. The potential measurement with stirring or unstirring and effects by double layer(DL) and triple layer (TL) will be discussed.