Diagnostics (Aug 2023)
Comparative Effectiveness of Two Models of Point-of-Care Ultrasound for Detection of Post-Void Residual Urine during Acute Ischemic Stroke: Preliminary Findings of Real-World Clinical Application
Abstract
We conducted a comparative study of two models of point-of-care ultrasound devices for measuring post-void residual urine (PVRU). We prospectively enrolled 55 stroke inpatients who underwent both real-time B-mode ultrasound (Device A) and automated three-dimensional (3D) scanning ultrasound (Device B), with a total of 108 measurements. The median PVRU volume of Device B was 40 mL larger than that of Device A. The PVRU difference between the devices was positively and linearly correlated with PVRU. The correlation of PVRU volume between the devices was strong, but the agreement level was only moderate. Measurement deviations were observed in 43 (40%) and 11 (10%) measurements with Device B and Device A, respectively. The PVRU volume was low in spherical bladder shapes but sequentially increased in triangular, undefined, ellipsoid, and cuboid bladder shapes. Further comparison of 60 sets of PVRU without measurement deviations revealed higher agreements between the devices at correction coefficients of 0.52, 0.66, and 0.81 for PVRU volumes of 200 mL, respectively. The automated 3D scanning ultrasound is more convenient for learning and scanning, but it exhibits larger measurement deviations. Real-time B-mode ultrasound accurately visualizes the urinary bladder but tends to underestimate the urinary bladder when the PVRU volume is large. Hence, real-time B-mode ultrasound with automated PVRU-based adjustment of calculation formulas may be a better solution for estimating bladder volume.