Diabetes & Metabolism Journal (Jan 2025)

Effectiveness of Predicted Low-Glucose Suspend Pump Technology in the Prevention of Hypoglycemia in People with Type 1 Diabetes Mellitus: Real-World Data Using DIA:CONN G8

  • Jee Hee Yoo,
  • Ji Yoon Kim,
  • Jae Hyeon Kim

DOI
https://doi.org/10.4093/dmj.2024.0039
Journal volume & issue
Vol. 49, no. 1
pp. 144 – 149

Abstract

Read online

We evaluated the effectiveness of the predictive low-glucose suspend (PLGS) algorithm in the DIA:CONN G8. Forty people with type 1 diabetes mellitus (T1DM) who used a DIA:CONN G8 for at least 2 months with prior experience using pumps without and with PLGS were retrospectively analyzed. The objective was to assess the changes in time spent in hypoglycemia (percent of time below range [%TBR]) before and after using PLGS. The mean age, sensor glucose levels, glucose threshold for suspension, and suspension time were 31.1±22.8 years, 159.7±23.2 mg/dL, 81.1±9.1 mg/dL, and 111.9±79.8 min/day, respectively. Overnight %TBR 0.05). Our findings support the PLGS in DIA:CONN G8 as an effective algorithm to improve night-time hypoglycemia and GRI in people with T1DM.

Keywords