Indian Pacing and Electrophysiology Journal (May 2023)

Pseudo loss of capture on 12 lead electrocardiogram in patient with an implantable cardiac defibrillator

  • Tuppence Richman,
  • Matthew Tung

Journal volume & issue
Vol. 23, no. 3
pp. 88 – 90

Abstract

Read online

This case report presents a 70-year-old-male, brought in by ambulance for assessment of chest pain and presyncope, with a paced ECG showing possible non-capture following ventricular pacing spikes. This was demonstrated to be pseudo non-capture with a 12-lead electrocardiogram performed in emergency and a device interrogation demonstrating ventricular capture.