Journal of Cardiothoracic Surgery (Jan 2025)

Remaining of contrast dye ice cap during PVI by cryoballoon ablation; a case report

  • Ahmad Yamini Sharif,
  • Entezar Mehrabi Nasab

DOI
https://doi.org/10.1186/s13019-024-03332-3
Journal volume & issue
Vol. 20, no. 1
pp. 1 – 4

Abstract

Read online

Abstract Pulmonary vein isolation (PVI) with cryoballoon (CB) ablation technology is widely used to treat drug-resistant atrial fibrillation (AF). During CB ablation, there is a possibility of forming an ice cap of contrast-color on top of the balloon. If automatic balloon deflate occurs before the ice cap dissolves, embolization to the systemic circulation is possible. This case report describes the remaining of a contrast-colored ice cap to the systemic circulation in the left superior pulmonary vein (LSPV) and the remaining contrast-colored ice cap throughout the balloon deflate and its gradual melting a few seconds after the balloon deflate in the right superior pulmonary vein (RSPV). Additionally, a strategy for its prevention is presented.