Arabian Journal of Chemistry (May 2017)
Spectrophotometric assay of creatinine in human serum sample
Abstract
A new spectrophotometric method for the analysis of creatinine concentration in human serum samples is developed. The method explores the oxidation of p-methylamino phenol sulfate (Metol) in the presence of copper sulfate and creatinine which yields an intense violet colored species with maximum absorbance at 530 nm. The calibration graph of creatinine by fixed time assay ranged from 4.4 to 620 μM. Recovery of creatinine in human serum samples varied from 101% to 106%. Limit of detection and limit of quantification were 0.145 μM and 0.487 μM respectively. Sandell’s sensitivity was 0.112 μg cm−2 and molar absorptivity was 0.101 × 104 L mol−1 cm−1. Within day precision was 2.5–4.8% and day-to-day precision range was 3.2–7.8%. The robustness and ruggedness of the method expressed in RSD values ranged from 0.78% to 2.12% and 1.32% to 3.46% respectively, suggesting that the developed method was rugged. This method provides good sensitivity and is comparable to standard Jaffe’s method with comparatively less interference from foreign substances.
Keywords