Arabian Journal of Chemistry (May 2017)
The comparison of partial least squares and principal component regression in simultaneous spectrophotometric determination of ascorbic acid, dopamine and uric acid in real samples
Abstract
Partial least squares (PLS1) and principal component regression (PCR) are two multivariate calibration methods that allow simultaneous determination of several analytes in spite of their overlapping spectra. In this research, a spectrophotometric method using PLS1 is proposed for the simultaneous determination of ascorbic acid (AA), dopamine (DA) and uric acid (UA). The linear concentration ranges for AA, DA and UA were 1.76–47.55, 0.57–22.76 and 1.68–28.58 (in μg mL−1), respectively. However, PLS1 and PCR were applied to design calibration set based on absorption spectra in the 250–320 nm range for 36 different mixtures of AA, DA and UA, in all cases, the PLS1 calibration method showed more quantitative prediction ability than PCR method. Cross validation method was used to select the optimum number of principal components (NPC). The NPC for AA, DA and UA was found to be 4 by PLS1 and 5, 12, 8 by PCR. Prediction error sum of squares (PRESS) of AA, DA and UA were 1.2461, 1.1144, 2.3104 for PLS1 and 11.0563, 1.3819, 4.0956 for PCR, respectively. Satisfactory results were achieved for the simultaneous determination of AA, DA and UA in some real samples such as human urine, serum and pharmaceutical formulations.