Effect of Baseline Impedance in Radiofrequency Delivery on Lesion Characteristics and the Relationship Between Impedance and Steam Pops
Abstract
ObjectiveTo explore the effects of baseline impedance (R) and power (P) on radiofrequency ablation (RFA) lesion characteristics and their correlation with steam pops using ThermoCool SmartTouch-SF (STSF) catheters in the porcine heart.MethodA porcine left ventricle was submerged in 37°C saline ex vivo, and the experiment was performed with various P (P = 30, 40, 50, and 60 W) and multiple R loads (R = 80–100, 100–140, 140–180, and 180–220 Ω) to reach the target ablation index (AI; AI = 350, 450, and 500) or reach the target ablation time using a fixed contact force (CF; CF = 10–15 g) and the same saline irrigation (30 W/8 ml/min or 40–60 W/15 ml/min), repeated five times under each condition.ResultsThe surface diameter, maximum diameter, depth, and volume of the lesions were strongly correlated with the AI (P = 40 W, R = 100–140 Ω, CF = 10–15 g) (r = 0.5412; r = 0.7889; r = 0.9366; and r = 0.913, respectively; all p < 0.05). As the value of R increased, the maximum diameter, depth, and volume of the lesions significantly increased (AI = 350, P = 30 W). Moreover, the higher the baseline value of R, the greater the absolute value of the R decrease (r = 0.9035, p < 0.05, Y = 0.2759 × X – 18.33). Under high power and high impedance, the occurrence rate of steam pops was high (P = 60 W, R = 180–220 Ω, AI when a steam pop occurred: 480 ± 26.5, ablation time: 11.29 ± 1.04 s).ConclusionRadiofrequency catheter ablation (RFCA) in power-controlled mode resulted in various lesion characteristics that were related to diverse baseline Rs. In addition, the incidence of steam pops was strongly correlated with high baseline R and high P.
Keywords