Veterinary World (Feb 2024)

Assessment of Vcheck® analyzer for rapid progesterone concentration measurement including recommendations for achieving the optimal breeding time in bitches

  • Supphathat Wutthiwitthayaphon,
  • Thanikran Suwannachote,
  • Saengtawan Arayatham,
  • Wisut Prasitsuwan,
  • Sakchai Ruenphet

DOI
https://doi.org/10.14202/vetworld.2024.427-433
Journal volume & issue
Vol. 17, no. 2
pp. 427 – 433

Abstract

Read online

Background and Aim: Serum progesterone concentration plays critical role in determining the optimal breeding time in bitches and diagnosing reproductive-related issues. This study aimed to conduct a comparative analysis of serum progesterone results obtained from commercial point-of-care immunological analyzers, namely, Vcheck®, with those obtained using chemiluminescent microparticle immunoassay (CMIA). Our overarching goal was to evaluate these analyzers’ accuracy and establish standardized guidelines for optimal breeding timing. Materials and Methods: Ninety-four serum samples from bitches were analyzed using the Vcheck® analyzer and compared with CMIA. Thorough documentation included the mean, standard deviation, 95% confidence interval (CI), and minimum and maximum values of serum progesterone concentrations. Furthermore, Pearson’s correlation coefficient, Lin’s concordance correlation coefficient, and the bias correction factor were meticulously recorded. Results: The mean progesterone concentration measured using the Vcheck® analyzer was significantly lower than that measured using CMIA, with a mean difference of 1.26 ng/mL of serum. The Bias correction factor was 0.935, which was nearly 1.00, indicating that the line of best-fit was on the perfect line of agreement, providing insight into the measurement accuracy. Pearson’s correlation coefficient, a measure of precision, was also close to 1 (0.939), confirming the reliability of the data. Furthermore, Lin’s concordance correlation coefficient was 0.877, indicating a fair overall agreement between the Vcheck® and CMIA methods. These results support the validity of the Vcheck® analyzer’s results. The present study was developed by aligning with established CMIA guidelines and adapting them using the range and 95% CI derived from each set of results, ensuring a standardized and rigorous approach. Conclusion: The Vcheck® analyzer provides a rapid assessment of serum progesterone concentration in bitches, with results comparable to those measured using the CMIA technique. However, when considering the use of the Vcheck® analyzer, it is recommended that the results should be interpreted carefully and the interpretation guidelines should be followed. In conclusion, Vcheck® provides a reliable and convenient method for veterinarian practitioners to measure canine progesterone levels in a clinical/hospital setting.

Keywords