IEEE Photonics Journal (Jan 2022)
Evaluation of Signal Degradation Due to Birefringence in a Multiple Reference Optical Coherence Tomography System With Polarization-Based Balanced Detection
Abstract
Although time-domain optical coherence tomography (TD-OCT) systems are straightforward to realize, the imaging speed, sensitivity, and imaging depth limit their range of applications. Multiple reference optical coherence tomography (MR-OCT) based on TD-OCT increases imaging range by about tenfold while providing sensitivity to image highly scattering biological samples. The multiple path-delays and free-space construction make MR-OCT also interesting for hybrid and compact systems, filling the gap between fibre-based and wafer-level integrated optical systems. We describe an optical configuration using a balanced detection scheme and the resulting signal properties due to the required use of polarizing optical components. We numerically simulate the signal properties using Jones calculus and compare the results with measurements. We discuss the origin of signal degradation due to birefringence of the sample in OCT and show that the quarter-wave plate in the sample arm of the Michelson interferometer can be adjusted to optimize the signal returning from a birefringent sample thereby improving the visibility of structures of interest. The theory discussed will be useful to understand and minimize signal degradation due to birefringence in Time-Domain and Fourier-Domain OCT systems.
Keywords