Diagnostics (Jan 2023)

Evaluation of InTray Cassettes Directly from Blood Cultures for the Diagnosis of Sepsis in Clinical Bacteriology Laboratories as an Alternative to Classic Culture Media

  • Alessandra Natale,
  • Saoussen Oueslati,
  • Alice Rochard,
  • Sien Ombelet,
  • Daniel Lopez-Baez,
  • Liselotte Hardy,
  • Jane Cunningham,
  • Céline Franquesa,
  • Olivier Vandenberg,
  • Jean-Baptiste Ronat,
  • Thierry Naas

DOI
https://doi.org/10.3390/diagnostics13030523
Journal volume & issue
Vol. 13, no. 3
p. 523

Abstract

Read online

Culture media is fundamental in clinical bacteriology for the detection and isolation of bacterial pathogens. However, in-house media preparation could be challenging in low-resource settings. InTray® cassettes (Biomed Diagnostics) could be a valid alternative as they are compact, ready-to-use media preparations. In this study, we evaluate the use of two InTray media as a subculture alternative for the diagnosis of bloodstream infections: the InTray® Müller-Hinton (MH) chocolate and the InTray® Colorex™ Screen. The InTray MH chocolate was evaluated in 2 steps: firstly, using simulated positive blood cultures (reference evaluation study), and secondly, using positive blood cultures from a routine clinical laboratory (clinical evaluation study). The Colorex Screen was tested using simulated poly-microbial blood cultures. The sensitivity and specificity of the InTray MH chocolate were respectively 99.2% and 90% in the reference evaluation study and 97.1% and 88.2% in the clinical evaluation study. The time to detection (TTD) was ≤20 h in most positive blood cultures (99.8% and 97% in the two studies, respectively). The InTray® MH Chocolate agar showed good performance when used directly from clinical blood cultures for single bacterial infections. However, mixed flora is more challenging to interpret on this media than on Colorex™ Screen, even for an experienced microbiologist.

Keywords