Scientific Reports (Oct 2020)
Evaluation of electrode-sample contact impedance under different curing humidity conditions during measurement of AC impedance of cement-based materials
Abstract
Abstract In this study, a simple method was proposed to calculate electrode-sample contact impedance in the cases of two-point and four-point measurements. The results indicated that when using the saturated calcium hydroxide solution (SCH) as conductive medium, the contact impedance in the four-point measurement is negligible for the impedance range of cement-based materials. The SCH can be used as a reference for correction of the contact impedance. A reasonable combination of curing humidity and different conductive media is recommended for the two-point measurement, which is suitable for testing the ACIS of cement-based materials. In a case of contact impedance not being precisely known, it is highly recommended that a four-point measurement with two different ratios of the length of the sample and the center spacing of the voltage electrodes (L/a) should be conducted to evaluate the effect of the contact impedance following the procedure proposed in this study.