Metrology and Measurement Systems (Mar 2022)

Period verification device for ventilation rate standard rods based on a digital thermal flowmeter

  • Jiacheng Hu,
  • Ying Sha,
  • Jinhui Cai,
  • Ying Liu,
  • Suijun Liu

DOI
https://doi.org/10.24425/mms.2022.138543
Journal volume & issue
Vol. vol. 29, no. No 1
pp. 231 – 244

Abstract

Read online

Ventilation rate is a physical index that is strictly controlled during cigarette manufacturing because an abnormal ventilation rate can affect the release of mainstream smoke, tar, and other components harmful to human health. Therefore, the standard rod is used for measuring the ventilation rate, which necessitates accurate and effective periodic inspections. In this study, we designed and built a set of special tobacco ventilation rate standard rods to assess the standard device during its verification period and used a digital thermal flowmeter as the flow standard.We determined the micro-pressure adjustment interval through fluid simulation, and conducted an experimental verification based on the simulation results. At the adjustment point where the differential pressure value was 0 Pa, the period verification device was tested under the standard values of 27.38%, 58.83%, and 71.95%. The results show that the measurement errors of the device are –0.42%, 0.55%, and –0.13% respectively, which all meet the verification regulation requirements and indicate that the device is applicable in practical situations.