EJNMMI Physics (May 2020)

Impact of dead time on quantitative 177Lu-SPECT (QSPECT) and kidney dosimetry during PRRT

  • Alessandro Desy,
  • Guillaume F. Bouvet,
  • Andrea Frezza,
  • Philippe Després,
  • Jean-Mathieu Beauregard

DOI
https://doi.org/10.1186/s40658-020-00303-0
Journal volume & issue
Vol. 7, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background Dead time may affect the accuracy of quantitative SPECT (QPSECT), and thus of dosimetry. The aim of this study was to quantify the effect of dead time on 177Lu-QSPECT and renal dosimetry following peptide receptor radionuclide therapy (PRRT) of neuroendocrine tumours. Methods QSPECT/CT was performed on days 1 and 3 during 564 personalized 177Lu-octreotate cycles in 166 patients. The dead-time data for each scanning time point was compiled. The impact of not correcting QSPECT for the dead time was assessed for the kidney dosimetry. This was also estimated for empiric PRRT by simulating in our cohort a regime of 7.4 GBq/cycle. Results The probability to observe a larger dead time increased with the injected activity. A dead-time loss greater than 5% affected 14.4% and 5.7% of QSPECT scans performed at days 1 and 3, respectively. This resulted in renal absorbed dose estimates that would have been underestimated by more than 5% in 5.7% of cycles if no dead-time correction was applied, with a maximum underestimation of 22.1%. In the simulated empiric regime, this potential dose underestimation would have been limited to 6.2%. Conclusion Dead-time correction improves the accuracy of dosimetry in 177Lu radionuclide therapy and is warranted in personalized PRRT.

Keywords