Urology Case Reports (Sep 2017)

Failure of the inflatable penile prosthesis due to abnormal folding of a low-profile reservoir – A selected case from an overall series and systematic review

  • Roberto Alejandro Navarrete,
  • Tao Cui,
  • Culley Carson,
  • Ryan Terlecki

DOI
https://doi.org/10.1016/j.eucr.2017.06.009
Journal volume & issue
Vol. 14, no. C
pp. 56 – 58

Abstract

Read online

We present a case from a running series of inflatable penile prosthesis failure due to improper folding of the Conceal™ reservoir. The Conceal™ Low-Profile reservoir gained popularity due to claims of improved cosmesis and ease of implantation. As the number of patients receiving this and other low-profile reservoirs increases, it is imperative to review and document any novel complications. While the Conceal™ reservoir may be preferred in ectopic placement, it may be more prone to fluid lockout facilitated by conformational change. Our review did not identify prior reports of improper folding, which we believe is unique to these low-profile reservoirs.

Keywords