Abstract
The measurement of cardiac troponin I (cTnI) is of vital importance for the early diagnosis of acute myocardial infarction. In this study, an enhanced electrochemiluminescent immunoassay for the highly sensitive and precise determination of cTnI was reported. A biomimetic chip with nepenthes peristome surface microstructures to achieve single-layer microbead arrays and integrated microelectrode arrays (MEAs) for ECL detection was microfabricated. Ru@SiO2 nanoparticles were prepared as signal amplificators labeling immunomagnetic beads. Dendrimer-encapsulated platinum nanoparticles (Pt DENs) were electrochemically modified on ITO MEAs. The resulting Pt DEN-modified ITO MEAs preserved good optical transparency and exhibited an approximately 20-fold ECL signal amplification compared to that obtained from bare ITO. The method made full use of the biomimetic chip with Pt DENs to develop single-layer immunomagnetic bead arrays with increasingly catalyzed electrochemical oxidation of the [Ru(bpy)3]2+–TPA system. Consequently, a limit of detection calculated as 0.38 pg/mL (S/N = 3) was obtained with excellent selectivity, demonstrating significant potential for the detection of cTnI in clinical diagnostics.