PLoS ONE (Jan 2019)

Intraocular pressure according to different types of tonometry (non-contact and Goldmann applanation) in patients with different degrees of bilateral tearing.

  • Bo Ram Seol,
  • Tae Gu Kang,
  • Bonhyeok Gu

DOI
https://doi.org/10.1371/journal.pone.0222652
Journal volume & issue
Vol. 14, no. 9
p. e0222652

Abstract

Read online

PurposeTo investigate intraocular pressure (IOP) readings by non-contact tonometry (NCT) and Goldmann applanation tonometry (GAT) for patients with different degrees of bilateral tearing.MethodsIn this study, we reviewed the medical charts of patients complaining of different degrees of bilateral tearing. The tear meniscus height (TMH) and IOP with NCT and GAT were measured. In each patient, a comparison of IOP readings between the eye with lower TMH and the contralateral eye with higher TMH was evaluated. The TMH was graded as follows: grade 1 (low): TMH ResultsA total of 120 eyes of 60 patients were enrolled. When comparing the two eyes of a patient, the eye with higher TMH showed higher NCT readings and larger difference in IOP readings between the two tonometries than the eye with lower TMH (P ConclusionEyes with higher TMH showed higher NCT readings and a larger difference in IOP between the two tonometries (NCT and GAT) than those with lower TMH. In patients with tearing, the NCT value may be inaccurate, so it is necessary to measure the GAT.