Brazilian Journal of Radiation Sciences (Jul 2017)

Filtration influence in a constant potential X-ray machine peak voltage measurements

  • Lucas Rodrigues Santos,
  • Vitor Vivolo,
  • Marcus Vinícius Teixeira Navarro,
  • Marcos Xavier,
  • Maria da Penha Albuquerque Potiens

DOI
https://doi.org/10.15392/bjrs.v5i2.261
Journal volume & issue
Vol. 5, no. 2

Abstract

Read online

This work shows the peak voltage measurements for several beam filtrations used in diagnostic radiology, using two types of non-invasive detectors; a voltage meter and a high-resolution spectrometer. The technique chosen for the voltage peak measurements with the spectrometer was the endpoint. The results were compared to the measured ones and showed good similarity to the nominal values. However the voltage meter detector used in this work presented errors for heavier filtrations.