Influence of Torque on Platform Deformity of the Tri-Channel Implant: Two- and Three-Dimensional Analysis Using Micro-Computed Tomography

  • Renata Costa de Morais,
  • Anselmo Agostinho Simionato,
  • Izabela Cristina Maurício Moris,
  • Graziela Bianchi Leoni,
  • Adriana Cláudia Lapria Faria,
  • Renata Cristina Silveira Rodrigues,
  • Ricardo Faria Ribeiro

DOI
https://doi.org/10.3390/medicina59071311
Journal volume & issue
Vol. 59, no. 7
p. 1311

Abstract

Read online

Background and Objectives: The insertion of the dental implant in the bone is an essential step in prosthetic rehabilitation. The insertion torque has the potential to distort the prosthetic platform, which can cause future biomechanical problems with the continuous action of occlusal forces. The aim of this study is to evaluate different insertion torques in the deformation of tri-channel platform connections through two- and three-dimensional measurements with micro-CT. Materials and Methods: A total of 164 implants were divided into groups (platform diameter and type): 3.5, 3.75, and 4.3 mm NP (narrow platform), and 4.3 mm RP (regular platform). Each implant–platform group was then divided into four subgroups (n = 10) with different torques: T45 (45 Ncm), T80 (80 Ncm), T120 (120 Ncm), and T150 (150 Ncm). The implant–abutment–screw assemblies were scanned and the images obtained were analyzed. Results: A significant difference was observed for the linear and volume measures between the different platforms (p p Conclusions: The 0.25 mm increment in the implant platform did not increase the resistance to the applied insertion torques; the 4.3 mm implant was significantly stronger compared to the 3.5 mm implant; and the proposed micro-CT analysis was considered valid for both the 2D and 3D analyses of micro-gaps, qualitatively and quantitatively.

Keywords