JACC: Asia (Sep 2021)

Application of Balloon-Expandable Transcatheter Heart Valve in Bicuspid Aortic Valve

  • Mi Chen, MD,
  • Jonathan Michel, MD,
  • Albert Markus Kasel, MD

Journal volume & issue
Vol. 1, no. 2
pp. 147 – 161

Abstract

Read online

Bicuspid aortic valve (BAV) remains challenging in transcatheter aortic valve replacement (TAVR) because of unfavorable anatomy. New-generation balloon-expandable valve (BEV) appears to be a valid alternative to surgery, especially in some Asian countries with a higher prevalence of BAV. This tutorial review summarizes current thinking about how to plan and implant BEV in BAVs using versatile techniques. First, the authors depict the main morphological characteristics of BAVs and their effects on the TAVR procedure. Next, the authors provide preprocedural analysis on sizing, obtaining the optimal deployment projection, and how to simplify valve-crossing. Finally, the authors provide step-by-step guidance on how to deploy the BEVs with evolved iterations in terms of specific anatomies, calcified annulus, and giant annulus.