Current Directions in Biomedical Engineering (Oct 2021)
Investigations on physico-chemical properties of TSPCU nonwoven for application as prosthetic venous valve
Abstract
Electrospinning is used for producing nonwovens for medical polymer-based implants, such as prosthetic valves or covered scaffolds. In this study, nonwovens for prosthetic venous valves are investigated regarding their morphology and mechanics in physiological medium. Spinning molds were developed based on previous venous valve leaflet designs, 3D printed in different sizes and covered with electrospun nonwovens. Samples were stored in a physiological 0.9% saline at 37°C to investigate the influence of fiber rearrangement and swelling in medium for several weeks. Two different nonwovens of thermoplastic silicone-based polycarbonaturethane (TSPCU) were compared. Tensile test results show that storage in medium has a relevant influence on the mechanical properties. SEM images of TSPCU show substantially increased fiber diameters after 8 days stored in medium. After detaching the valve leaflet nonwovens from the molds, shrinkage of the material of approximately 12% was detected. A suitable valve size could be identified for joining with the stent structure into an interventional prosthetic venous valve. The results demonstrate the influence of storage conditions on the morphological and mechanical properties of electrospun TSPCU nonwovens. For development and dimensioning of venous valve leaflets, this change in mechanical behavior and possible shrinkage of the material has to be considered.
Keywords