Open Chemistry (Jan 2016)

Development and validation of a new spectrofluorimetric method for the determination of some beta-blockers through fluorescence quenching of eosin Y. Application to content uniformity test

  • Derayea Sayed M,
  • Omar Mahmoud A,
  • Abdel-Lateef Mohamed Aboel-Kasem,
  • Hassan Ahmed I.

DOI
https://doi.org/10.1515/chem-2016-0024
Journal volume & issue
Vol. 14, no. 1
pp. 258 – 266

Abstract

Read online

A simple, rapid, sensitive and economic spectrofluorimetric method has been developed and validated for determination of some β-adrenergic blocking agents namely; betaxolol hydrochloride (BTX), carvedilol (CAR), labetalol hydrochloride (LBT), nebivolol hydrochloride (NEB) and propranolol hydrochloride (PRO). The method is based on the quenching effect of the cited drugs on the fluorescence intensity of eosin Y at pH 3.4 (acetate buffer). The fluorescence quenching is due to the formation of an ion-pair complex and was measured without extraction at 545 nm (λex. 301.5 nm). The factors affecting the formation of the ion-pair complex were carefully studied and optimized. Under the optimal conditions, the linear ranges for the relationship between the fluorescence quenching value and the concentration of the investigated drugs were 100-2500, 150-2500 and 50-2250 ng mL-1 for (BTX, CAR), (LBT, NEB) and (PRO) respectively. The method was validated according to ICH guidelines and was applied for determination of the cited drugs in pharmaceutical dosage forms with excellent recoveries. In addition, content uniformity testing of some commercial dosage forms was checked by the proposed method.

Keywords