Physics and Imaging in Radiation Oncology (Jul 2018)

An evaluation of systematic errors on marker-based registration of computed tomography and magnetic resonance images of the liver

  • Thomas Woolcot,
  • Evanthia Kousi,
  • Emma Wells,
  • Katharine Aitken,
  • Helen Taylor,
  • Maria A. Schmidt

Journal volume & issue
Vol. 7
pp. 27 – 31

Abstract

Read online

We demonstrated a general method to evaluate systematic errors related to Magnetic Resonance (MR) imaging sequences in marker-based co-registration of MR and Computed Tomography (CT) images, and investigated the effect of MR image quality in the co-registration process using clinical MR and CT protocols for stereotactic ablative body radiotherapy (SABR) planning of the liver. Small systematic errors (under 1.6 mm) were detected, unlikely to be a clinical risk to liver SABR. The least favourable marker configuration was found to be a co-planar arrangement parallel to the transaxial image plane. Keywords: Registration, Fiducial markers, Stereotactic radiotherapy, Liver