International Journal of Metrology and Quality Engineering (Jan 2011)

Uncertainty evaluation in the calibration of Pt/Pd thermocouples up to copper freezing point at NIS-Egypt

  • Megahed F.M.,
  • Abdelaziz Y.A.

DOI
https://doi.org/10.1051/ijmqe/2011003
Journal volume & issue
Vol. 2, no. 1
pp. 13 – 17

Abstract

Read online

The dissemination of traceability to the International Temperature Scale of 1990 (ITS-90) through comparison calibration of thermocouples, is playing an important role in metrology institutes, secondary laboratories and industrial measurements. The previous studies of the Pt/Pd thermocouples at NIS-Egypt, shows promise of achieving significant improvements uncertainty of temperature measurements up to 960 °C. In this paper we describe the uncertainty assessment on the measurements of the calibration of Pt/Pd thermocouples at fixed point cells up to the freezing point of copper and by the comparison technique following a defined heat treatment at 1100 °C. All the measurements on Pt/Pd thermocouples were carried out using NIS facilities.

Keywords