PLoS ONE (Jan 2020)

Hemoglobin point-of-care testing in rural Gambia: Comparing accuracy of HemoCue and Aptus with an automated hematology analyzer.

  • Stefan A Nass,
  • Ilias Hossain,
  • Chilel Sanyang,
  • Bakary Baldeh,
  • Dora I A Pereira

DOI
https://doi.org/10.1371/journal.pone.0239931
Journal volume & issue
Vol. 15, no. 10
p. e0239931

Abstract

Read online

BackgroundAnemia is one of the most impactful nutrient deficiencies in the world and disproportionately affects children in low-resource settings. Point-of-care devices (PoCDs) measuring blood hemoglobin (Hb) are widely used in such settings to screen for anemia due to their low cost, speed, and convenience. Here we present the first iteration of Aptus, a new PoCD which measures Hb and hematocrit (HCT).AimTo evaluate the accuracy of Aptus and HemoCue® Hb 301 against an automated hematology analyzer (Medonic®) in Gambian children aged 6-35 months and the Aptus' usage in the field.MethodsAptus, HemoCue® and Medonic® were compared using venous blood (n = 180), and Aptus and HemoCue® additionally using capillary blood (n = 506). Agreement was estimated using Bland-Altman analysis and Lin's concordance. Usage was assessed by error occurrence and user experience.ResultsMean Hb values in venous blood did not significantly differ between Aptus and HemoCue® (10.44±1.05 vs 10.56±0.93g/dl, p>0.05), but both measured higher Hb concentrations than Medonic® (9.75±0.99g/dl, pConclusionBoth PoCDs showed a relatively limited bias but large LoA. Aptus and HemoCue® showed similar accuracy, while both overestimated Hb levels. Aptus showed promise, with its operation unimpaired by field conditions as well as being able to show HCT values.