Cogent Engineering (Jan 2019)
The effect of extruder pressure of 3D bioprinter on hardness and compressive of scaffold bovine hydroxyapatite
Abstract
The aims of the study are to find out bio-printer extruder pressure effect toward the hardness and compressive strength of bovine hydroxyapatite (BHA). BHA and Glycerin are mixed with a 1:1 ratio. After 24 h bioink will be formed and can be used as a material to print scaffold. The scaffold is printed with a pressure variation of 100, 150, 200 and 250 KPa. After the printing process is done, the scaffold is kept on heat bed for 10 min with 100°C temperature. The hardness test result in this study has the highest value of 24.23 ± 1.22 HVN and 2.03 ± 0.12 HVN for its lowest. The compressive strength test has the highest value of 5.71 ± 0.42 MPa with the lowest value of 3.22 ± 0.13 MPa. The observation of BHA scaffold pores using a microscope shows that the pores have an average size of 100–200 µm.