Nuclear Engineering and Technology (Dec 2023)

True coincidence summing correction factor for point source geometry with PHITS

  • Esra Uyar

Journal volume & issue
Vol. 55, no. 12
pp. 4472 – 4476

Abstract

Read online

In this study, it has been shown that the true coincidence summing correction factor can be obtained for the first time using the PHITS Monte Carlo program. Determining this correction factor using different methods and tools in each laboratory to increase the possibility of achieving high-efficiency measurement conditions is still popular in gamma-ray spectrometry. By using 133Ba, 152Eu, 134Cs, and 60Co point sources, the true coincidence summing factor was investigated in both near and far counting geometries for 15 different energy values. GESPECOR software was used to validate the results obtained with PHITS. A remarkable agreement was obtained between PHITS and GESPECOR, with a discrepancy of less than 3%. With this study, a new tool has been proposed to obtain the true coincidence summing factor, which is one of the significant correction factors investigated/calculated in gamma-ray spectrometric studies.

Keywords