BioTechniques (Sep 2012)

Improving the limit of detection for Sanger sequencing: A comparison of methodologies for KRAS variant detection

  • Colin J. Davidson,
  • Emily Zeringer,
  • Kristen J. Champion,
  • Marie-Pierre Gauthier,
  • Fawn Wang,
  • Jerry Boonyaratanakornkit,
  • Julie R. Jones,
  • Edgar Schreiber

DOI
https://doi.org/10.2144/000113913
Journal volume & issue
Vol. 53, no. 3
pp. 182 – 188

Abstract

Read online

Fluorescent dye terminator Sanger sequencing (FTSS), with detection by automated capillary electrophoresis (CE), has long been regarded as the gold standard for variant detection. However, software analysis and base-calling algorithms used to detect mutations were largely optimized for resequencing applications in which different alleles were expected as heterozygous mixtures of 50%. Increasingly, the requirements for variant detection are an analytic sensitivity for minor alleles of A transitions and from 50%–5% to 5% for G>C and G>T transversions. In addition, we use two different sample types to compare the limit of detection of sequence variants in codons 12 and 13 of the KRAS gene between Sanger sequencing and other methodologies including shifted termination assay (STA) detection, single-base extension (SBE), pyrosequencing (PS), high- resolution melt (HRM), and real-time PCR (qPCR).

Keywords