Российский технологический журнал (Apr 2016)

LENS ACOUSTIC MICROSCOPE WITH A LINEAR ARRAY FOR QUANTITATIVE CHARACTERIZATION OF LAYERED OBJECTS

  • S. A. Titov,
  • R. G. Maev,
  • A. N. Bogachenkov

DOI
https://doi.org/10.32362/2500-316X-2016-4-2-25-30
Journal volume & issue
Vol. 4, no. 2
pp. 25 – 30

Abstract

Read online

In this paper, the acoustic microscope with a linear ultrasonic array and a cylindrical lens has been proposed in order to carry out quantitative characterization of layered regions of the sample under research. The determination method based on recording the microscope array spatio-temporal signal of the layer thickness and velocities of the longitudinal and transverse waves in the layer material has been developed. By using a test sample, it was shown that the error of measurement is about 1 % for both the thickness and the velocity of the longitudinal wave, and it is 3 % for the velocity of the transverse wave.