BMJ Open Ophthalmology (Apr 2024)

Red filter meibography by smartphones in patients with meibomian gland dysfunction: a validity and reliability study

  • Rina La Distia Nora,
  • Ratna Sitompul,
  • Aria Kekalih,
  • Gisela Haza Anissa,
  • Syska Widyawati,
  • Prasandhya Astagiri Yusuf

DOI
https://doi.org/10.1136/bmjophth-2023-001266
Journal volume & issue
Vol. 9, no. 1

Abstract

Read online

Objective The objective of this study is to determine the validity and reliability of the red filter meibography by smartphone compared with infrared in assessing meibomian gland drop-out.Methods and analysis An analytical cross-sectional study was done with a total of 35 subjects (68 eyes) with suspected MGD based on symptoms and lid morphological abnormalities. Meibomian glands were photographed using two smartphones (Samsung S9 and iPhone XR) on a slit-lamp with added red filter. Images were assessed subjectively using meiboscore by the two raters and drop-out percentages were assessed by ImageJ.Results There was no agreement in meiboscore and a minimal level of agreement in drop-out percentages between red filter meibography and infrared. Inter-rater reliability showed no agreement between two raters. Intra-rater reliability demonstrated weak agreement in rater 1 and no agreement in rater 2.Conclusion Validity of the red filter meibography technique by smartphones is not yet satisfactory in evaluating drop-out. Further improvement on qualities of images must be done and research on subjective assessment was deemed necessary due to poor results of intrarater and inter-rater reliability.