Animals (Dec 2023)

Comparison of the Automated OvaCyte Telenostic Faecal Analyser versus the McMaster and Mini-FLOTAC Techniques in the Estimation of Helminth Faecal Egg Counts in Equine

  • Nagwa Elghryani,
  • Conor McAloon,
  • Craig Mincher,
  • Trish McOwan,
  • Theo de Waal

DOI
https://doi.org/10.3390/ani13243874
Journal volume & issue
Vol. 13, no. 24
p. 3874

Abstract

Read online

Gastrointestinal helminth parasites continue to be a significant threat to the health of equine. OvaCyte Telenostic (OCT) (Telenostic Ltd., Kilkenny, Ireland) has developed an automated digital microscope utilising Artificial Intelligence to identify and count the clinically important helminth species in equine, bovine, and ovine host species. In this paper, the performance of the OCT analyser was evaluated for the detection and counting of equine helminth species parasites and its performance compared to the currently accepted benchmark methods of faecal egg counts being the McMaster and the Mini-FLOTAC techniques. A pairwise comparison of tests was assessed based on the correlation of egg counts and Cohen’s kappa agreement statistics for dichotomized outcomes. Bayesian latent class analysis was used to estimate the sensitivity and specificity of all three techniques in the absence of a gold standard for four helminth parasites (strongyles, Anoplocephala spp., Parascaris spp. and Strongyloides westeri). Based on the analysis of 783 equine faecal samples, we found a high level (ρ ≥ 0.94) of correlation between each pairwise comparison of techniques for strongyle egg counts. Cohen’s kappa agreement between techniques was high for strongyles and S. westeri, moderate for Parascaris spp., and low for Anoplocephala spp. All three techniques had a high sensitivity and specificity (>0.90) for strongyles. Across helminth parasites, the sensitivity of the OCT was the highest of the three techniques evaluated for strongyles (0.98 v 0.96 and 0.94), Anoplocephala spp. (0.86 v 0.44 and 0.46) and Parascaris spp. (0.96 v 0.83 and 0.96); but lowest for S. westeri (0.74 v 0.88 and 0.88), compared to McMaster and Mini-FLOTAC, respectively. In terms of specificity, OCT was the lowest in two species (Parascaris spp. 0.96, Anoplocephala spp. 0.95). In conclusion, OCT has a sensitivity and specificity statistically similar to both McMaster and Mini-FLOTAC, and had a higher correlation with Mini-FLOTAC. The OCT point of care faecal analyser offers improved workflow, test turn-around time and does not require trained laboratory personnel to operate or interpret the results

Keywords