Journal of Isotopes (Nov 2016)

Development of a Grid Ionization Chamber for a Dram of α Ray

  • YANG Lu;WANG Qiang;ZHENG Yu-lai;WANG Guo-bao

DOI
https://doi.org/10.7538/tws.2016.29.04.0230
Journal volume & issue
Vol. 29, no. 4
pp. 230 – 234

Abstract

Read online

This article introduced the parallel grid ionization chamber that used to measure the α radioactivity, which had a independent vacuum system. The system was composed of main body of the chamber, gas-filled and electronics system. Energy resolution was 26 keV for 239Pu, background was 10 counts for one hour from 4 MeV to 6 MeV energy range, the stability of 24 hours was less than 0.5% . The chamber can measure the energy of nuclide and analyze the energy spectrum structure to identify nuclear.