Heliyon (Oct 2018)

Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams

  • Nashrulhaq Tagiling,
  • Raizulnasuha Ab Rashid,
  • Siti Nur Amirah Azhan,
  • Norhayati Dollah,
  • Moshi Geso,
  • Wan Nordiana Rahman

Journal volume & issue
Vol. 4, no. 10
p. e00864

Abstract

Read online

Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson® Expression® 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry.

Keywords