Nuclear Technology and Radiation Protection (Jan 2020)

Efficiency calibration of a well-type HPGe detector using experimental and Monte Carlo simulation techniques

  • Dalaka Ekaterini,
  • Kuburas Georgios,
  • Eleftheriadis Konstantinos,
  • Anagnostakis Marios J.

DOI
https://doi.org/10.2298/NTRP2002121D
Journal volume & issue
Vol. 35, no. 2
pp. 121 – 129

Abstract

Read online

Well-type high-purity germanium detectors are well suited for the analysis of small samples, as they combine high detection efficiency with low background radiation. The well geometry however makes efficiency calibration more difficult than that of ordinary HPGe detectors, due to intense true coincidence and possibly random summing effects. Such a detector has been installed at the Environmental Radioactivity Laboratory of the National Centre for Scientific Research "Demokritos". For the calibration of this detector, experimental and Monte Carlo simulation techniques were applied. To this end, calibration sources were produced from the radionuclides available at the Environmental Radioactivity Laboratory. Starting from the geometrical characteristics of the detector as provided by the manufacturer, using the calibration sources and applying Monte Carlo simulation techniques, the detector was characterized and peak efficiency, as well as total-to-peak calibration curves were produced. The results of the calibration finally obtained by simulation are found to be in good agreement with the respective experimental calibration results.