Sensors (Jun 2008)

Simultaneous Chronoamperometric Sensing of Ascorbic Acid and Acetaminophen at a Boron-Doped Diamond Electrode

  • Ciprian Radovan,
  • CodruÅ£a Cofan

Journal volume & issue
Vol. 8, no. 6
pp. 3952 – 3969

Abstract

Read online

Cyclic voltammetry (CV) and chronoamperometry (CA) have been used to sense and determine simultaneously L-ascorbic acid (AA) and acetaminophen (AC) at a boron-doped diamond electrode (BDDE) in a Britton-Robinson buffer solution. The calibration plots of anodic current peak versus concentration obtained from CV and CA data for both investigated compounds in single and di-component solutions over the concentration range 0.01 mM – 0.1 mM proved to be linear, with very good correlation parameters. Sensitivity values and RSD of 2-3% were obtained for various situations, involving both individual and simultaneous presence of AA and AC. The chronoamperometric technique associated with standard addition in sequential one step and/or two successive and continuous chronoamperograms at two characteristic potential levels represented a feasible option for the simultaneous determination of AA and AC in real sample systems such as pharmaceutical formulations. The average values indicated by the supplier were confirmed to a very close approximation from chronoamperomgrams by using several additions with the application of suitable current correction factors.