PLoS ONE (Jan 2022)

Power output from 12 brands of contemporary LED light-curing units measured using 2 brands of radiometers.

  • Cristiane Maucoski,
  • Richard B Price,
  • Cesar A Arrais,
  • Braden Sullivan

DOI
https://doi.org/10.1371/journal.pone.0267359
Journal volume & issue
Vol. 17, no. 7
p. e0267359

Abstract

Read online

BackgroundGiven the increasing use of photo-activated resins in dentistry, dentists and researchers need a user-friendly dental radiometer to measure the power output from dental light-curing units (LCUs).ObjectiveOur goal was to measure the accuracy of two brands of dental radiometers in reporting the power (mW) from twelve brands of contemporary LCUs compared to a 'gold standard' (GS) reference value obtained from an integrating sphere attached to a fiberoptic spectroradiometer.MethodsThe power output was measured from two units of 12 brands of LCUs, five times on the ''GS" system, five times on two Bluephase Meter II dental radiometers, and five times on two Mini Gig hand-held spectroradiometers. The emission spectrum was also recorded using the 'GS' integrating sphere. The power values reported by each meter were subjected to t-tests to compare the two examples of each LCU, and 3-way ANOVA followed by Bonferroni's post-hoc tests. Regression analyses were also performed to determine the relationship between the data from the hand-held radiometers and the 'GS' integrating sphere.ResultsThere was a large difference in the power values (mW) and the emission spectra from the 12 brands of LCUs on their standard-settings (p 98%).ConclusionWe concluded that the power values reported from both brands of dental radiometers we tested were accurate, provided that the light source did not emit wavelengths of light that were beyond the radiometer's detection limit.