Force-degradation pattern of six different orthodontic elastomeric chains.

  • Ah Mirhashemi,
  • A Saffarshahroudi,
  • A Sodagar,
  • M Atai

Journal volume & issue
Vol. 9, no. 4

Abstract

Read online

An ideal orthodontic force system should exert continuous light force. Thus, many efforts have been made to improve the memory characteristics of elastomeric chains. The aim of this study was to compare elastomeric chains (ECs) claimed by their manufacturers to offer high memory with traditional ones according to their force-extension diagrams.In this in-vitro study, ECs were divided into six groups, each containing 40 pieces of chain, from three brands (American Orthodontics, GAC and Ortho-Technology). Each brand was divided into two groups with respect to their claimed characteristics (with or without memory). Each sample was stretched to twice its original length and kept constant in 37°C distilled water. Force-extension diagrams were drawn by universal testing machine at 0,1,8,24,72 hours and 1, 2, 4-week intervals. Additionally, the amounts of elongation required to deliver 200 g force were calculated. To compare the results, ANOVA and Tukey tests were performed.Force-decay rate was significantly different between traditional and memory chains (p