PLoS ONE (Jan 2021)
Impact of image averaging on vessel detection using optical coherence tomography angiography in eyes with macular oedema and in healthy eyes
Abstract
Purpose To assess the repeatability of multiple automatic vessel density (VD) measurements and the effect of image averaging on vessel detection by optical coherence tomography angiography (OCTA). Methods An observational study was conducted in a series of healthy volunteers and patients with macular oedema. Five sequential OCTA images were acquired for each eye using the OptoVue HD device. The effect of the averaging of the 5 acquisitions on vessel detection was analysed quantitatively using a pixel-by-pixel automated analysis. In addition, two independent retina experts qualitatively assessed the change in vessel detection in averaged images segmented in 9 boxes and compared to the first non-averaged image. Results The automatic VD measurement in OCTA images showed a good repeatability with an overall mean intra-class correlation coefficient (ICC) of 0.924. The mean ICC was higher in healthy eyes compared to eyes with macular oedema (0.877 versus 0.960; p Conclusion The automatic VD measurement using the OptoVue HD device showed a good repeatability in 5 acquisitions in a row setting. Averaging images increased vessel detection, and in about a third of boxes, decreased the background noise, both in healthy eyes and, in a greater proportion, in eyes with macular oedema.