Prosthesis (Jul 2023)

Evaluation of Rotational Stability and Stress Shielding of a Stem Optimized for Hip Replacements—A Finite Element Study

  • Mario Ceddia,
  • Bartolomeo Trentadue

DOI
https://doi.org/10.3390/prosthesis5030048
Journal volume & issue
Vol. 5, no. 3
pp. 678 – 693

Abstract

Read online

The natural distribution of stress in the femur is altered when total hip arthroplasty (THA) is performed. In fact, when a stem is inserted inside the femur, there is a variation in stress due to the difference in rigidity between the material with which the stem is made and the femur. This generates the phenomenon of stress shielding. The aim of this study is to design an optimized prosthesis that guarantees an excellent rotational stability and a reduced stress shielding. Methods: Through the finite element method (FEM), the mechanical behavior of the stem subjected to the loads described by ISO 7206-4:2010 is studied. Results: Through topological optimization, there is a reduction in stress shielding in the proximal zone of 31.46%. The addition of ridges on the dorsal side of the stem also improves rotational stability by 27.82%. Conclusions: The decrease in stiffness that is recorded with the optimized stem guarantees a greater distribution of stress on the bone. The presence of dorsal ridges also favors the corticalization of the bone as it loads the bone near the dorsal, ensuring further stability. The perforated prosthesis presented in this study shows an increase in primary stability and an improvement in rotational stability as there is also a bone regrowth inside the prosthesis.

Keywords