International Journal of Hyperthermia (Jan 2021)

The efficacy and safety of MR-HIFU and US-HIFU in treating uterine fibroids with the volume <300 cm3: a meta-analysis

  • Liang Yu,
  • Shu Zhu,
  • Huiyuan Zhang,
  • Anqi Wang,
  • Guodong Sun,
  • JiaLe Liang,
  • Xiuli Wang

DOI
https://doi.org/10.1080/02656736.2021.1954245
Journal volume & issue
Vol. 38, no. 1
pp. 1126 – 1132

Abstract

Read online

Background High-intensity focused ultrasound (HIFU) is a promising and non-invasive therapy for symptomatic uterine fibroids. Currently, the main image-guided methods for HIFU include magnetic resonance-guided (MR-HIFU) and ultrasound-guided (US-HIFU). However, there are few comparative studies on the therapeutic efficacy and safety of MR-HIFU and US-HIFU in treating symptomatic uterine fibroids with a volume <300 cm3. Objective We performed this meta-analysis to evaluate the efficacy and safety of MR-HIFU and US-HIFU in treating symptomatic uterine fibroids with a volume <300 cm3. Methods We searched relevant literature in PubMed, EMBASE, Cochrane Library CNKI from inception until 2021. The mean value, the proportion, and their 95% confidence intervals (CIs) were measured by random-effects models. Publication bias was assessed using funnel plots. Results 48 studies met our inclusion criteria—28 describing MR-HIFU and 20 describing US-HIFU. The mean non-perfused volume rate (NPVR) was 81.07% in the US-HIFU group and 58.92% in the MR-HIFU group, respectively. The mean volume reduction rates at month-3, month-6, and month-12 were 42.42, 58.72, and 65.55% in the US-HIFU group, while 34.79, 37.39, and 36.44% in the MR-HIFU group. The incidence of post-operative abdominal pain and abnormal vaginal discharge in the US-HIFU group was lower than that of MRI-HIFU. However, post-operative skin burn and sciatic nerve pain were more common in the US-HIFU group compared with MRI-HIFU. The one-year reintervention rate after MR-HIFU was 13.4%, which was higher than 5.2% in the US-HIFU group. Conclusion US-HIFU may show better efficiency and safety than MR-HIFU in treating symptomatic fibroids with a volume <300 cm3.

Keywords