Clinical Ophthalmology (Mar 2015)

Corneal decompensation following filtering surgery with the Ex-PRESS® mini glaucoma shunt device

  • Tojo N,
  • Hayashi A,
  • Miyakoshi A

Journal volume & issue
Vol. 2015, no. default
pp. 499 – 502

Abstract

Read online

Naoki Tojo, Atsushi Hayashi, Akio Miyakoshi Department of Ophthalmology, Graduate School of Medicine and Pharmaceutical Sciences, University of Toyama, Toyama, Japan Purpose: To report a case of corneal decompensation due to the Ex-PRESS® mini glaucoma shunt device (Ex-PRESS).Patient and methods: A 75-year-old man had pseudoexfoliation glaucoma in his right eye. He underwent filtration surgery with Ex-PRESS. His intraocular pressure was 7 mmHg after 9 months.Results: We observed partial decompensation of the corneal endothelium adjacent to the filtering bleb. Specular microscopy revealed a marked decrease in the endothelial cell density at the center of the cornea.Conclusion: Anterior segment optical coherence tomography is very useful for evaluating corneal edema and the position of Ex-PRESS. It is important to follow up with an examination of the corneal endothelial cells. Keywords: Ex-PRESS, bullous keratopathy, trabeculectomy, complication, cornea