Photoacoustics (Dec 2022)

Posterior photoacoustic/ultrasound imaging of the periodontal pocket with a compact intraoral transducer

  • Lei Fu,
  • Reza Khazaeinezhad,
  • Ali Hariri,
  • Baiyan Qi,
  • Casey Chen,
  • Jesse V. Jokerst

Journal volume & issue
Vol. 28
p. 100408

Abstract

Read online

Periodontitis is a public issue and imaging periodontal pocket is important to evaluate periodontitis. Regular linear transducers have limitations in imaging the posterior teeth due to their geometry restrictions. Here we characterized a transducer that can image the posterior teeth including assessment of periodontal pockets via a combination of photoacoustic and ultrasound imaging. Unlike conventional transducer design, this device has a toothbrush-shaped form factor with a side-view transducer to image molars (total size: 1 ×1.9 cm). A laser diode was integrated as the light source to reduce the cost and size and facilitates clinical transition. The in vivo imaging of a molar of a periodontal patient demonstrated that the transducer could image in the posterior area of gum in vivo; the value determined by imaging was within 7 % of the value measured clinically.