Pharmaceuticals (Mar 2024)

Development of Analytical Method for the Quantitation of Monoclonal Antibodies Solutions via Raman Spectroscopy: The Case of Bevacizumab

  • Michail Lykouras,
  • Panagiota Papaspyridakou,
  • Olga E. Makri,
  • Constantine D. Georgakopoulos,
  • Malvina G. Orkoula

DOI
https://doi.org/10.3390/ph17040446
Journal volume & issue
Vol. 17, no. 4
p. 446

Abstract

Read online

Personalized dosages of monoclonal antibodies are being used more regularly to treat various diseases, rendering their quantitation more essential than ever for the right dose administration to the patients. A promising alternative, which overcomes the obstacles of the well-established chromatographic techniques regarding the quantification of biopharmaceuticals, is Raman spectroscopy. This study aimed to develop and validate a novel analytical method for the quantitation of bevacizumab in solutions via Raman spectroscopy. For this purpose, a droplet of the solution was left to dry on a highly reflective carrier and a home-made apparatus was employed for rotation of the sample. Hence, each recorded Raman spectrum was the average of the signal acquired simultaneously from multiple points on a circular circumference. The method was validated, and the detection limit of the antibody was found to be 1.06 mg/mL. Bevacizumab was found to be highly distributed at the formed coffee ring of the dried droplet, though this was a function of solution concentration. Finally, Raman spectra at different distances on the coffee ring were obtained from the four quarters. The lowest bevacizumab detection limit was found at a distance of 75 μm from the external side of the coffee ring and it was determined to be equal to 0.53 mg/mL.