Critical Care Explorations (Jul 2024)

Quantitative Comparison of Ventilation Parameters of Different Approaches to Ventilator Splitting and Multiplexing

  • Doowon Kim, MSc,
  • Steven Roy, MDCM, FRCPC,
  • Paul McBeth, MD, MASc, FRCSC,
  • Jihyun Lee, PhD

DOI
https://doi.org/10.1097/CCE.0000000000001113
Journal volume & issue
Vol. 6, no. 7
p. e1113

Abstract

Read online

CONTEXT:. Amid the COVID-19 pandemic, this study delves into ventilator shortages, exploring simple split ventilation (SSV), simple differential ventilation (SDV), and differential multiventilation (DMV). The knowledge gap centers on understanding their performance and safety implications. HYPOTHESIS:. Our hypothesis posits that SSV, SDV, and DMV offer solutions to the ventilator crisis. Rigorous testing was anticipated to unveil advantages and limitations, aiding the development of effective ventilation approaches. METHODS AND MODELS:. Using a specialized test bed, SSV, SDV, and DMV were compared. Simulated lungs in a controlled setting facilitated measurements with sensors. Statistical analysis honed in on parameters like peak inspiratory pressure (PIP) and positive end-expiratory pressure. RESULTS:. Setting target PIP at 15 cm H2O for lung 1 and 12.5 cm H2O for lung 2, SSV revealed a PIP of 15.67 ± 0.2 cm H2O for both lungs, with tidal volume (Vt) at 152.9 ± 9 mL. In SDV, lung 1 had a PIP of 25.69 ± 0.2 cm H2O, lung 2 at 24.73 ± 0.2 cm H2O, and Vts of 464.3 ± 0.9 mL and 453.1 ± 10 mL, respectively. DMV trials showed lung 1’s PIP at 13.97 ± 0.06 cm H2O, lung 2 at 12.30 ± 0.04 cm H2O, with Vts of 125.8 ± 0.004 mL and 104.4 ± 0.003 mL, respectively. INTERPRETATION AND CONCLUSIONS:. This study enriches understanding of ventilator sharing strategy, emphasizing the need for careful selection. DMV, offering individualization while maintaining circuit continuity, stands out. Findings lay the foundation for robust multiplexing strategies, enhancing ventilator management in crises.