Journal of Arrhythmia (Dec 2017)

Pacing failure caused by automatic pacing threshold adjustment system

  • Jun Kishihara,
  • Shinichi Niwano, MD,
  • Hidehira Fukaya,
  • Ryo Nishinarita,
  • Ai Horiguchi,
  • Hironori Nakamura,
  • Tazuru Igarashi,
  • Naruya Ishizue,
  • Tamami Fujiishi,
  • Tomoharu Yoshizawa,
  • Jun Oikawa,
  • Akira Satoh,
  • Masami Murakami,
  • Junya Ako

Journal volume & issue
Vol. 33, no. 6
pp. 637 – 639

Abstract

Read online

Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient. Keywords: Automatic pacing threshold adjustment algorithm, Pacing failure