Kaohsiung Journal of Medical Sciences (Sep 2008)

Catheter Migration After Implantationan Intrathecal Baclofen Infusion Pump for Severe Spasticity: A Case Report

  • Tung-Chou Li,
  • Mei-Hsiu Chen,
  • Jing-Shan Huang,
  • Jun-Yeen Chan,
  • Yaun-Kai Liu,
  • Ming-Hong Chen

DOI
https://doi.org/10.1016/S1607-551X(09)70007-5
Journal volume & issue
Vol. 24, no. 9
pp. 492 – 497

Abstract

Read online

We report a case of intrathecal baclofen infusion pump implantation complicated by migration of the catheter tip. A 55-year-old man required an intrathecal baclofen infusion for severe spasticity 4 years after a cervical spinal cord injury with incomplete tetraparesis. Twelve months after initial implantation of the device, the patient began to experience a recurrence of trunk tightness and spasticity. Subsequent X-ray and computed tomography evaluations of the catheter system revealed pooling of contrast medium outside of the intrathecal distribution in the lumbar subcutaneous region of the back and therefore migration of the pump catheter tip. At surgical revision, emphasis was placed on minimizing the length of catheter outside of the spine and securing the catheter in the supraspinous fascia with a right-angled anchor. The distance between the anchors and the entry point of the catheter into the supraspinous fascia was also reduced to prevent slipping when the patient bends forward. After surgery, the patient's spasticity improved and, 1 year later, he has experienced no further complications during follow-up, requiring an average baclofen dose of 150 mg/day. Here, we describe several surgical methods intended to secure the intrathecal catheter and prevent catheter migration. Other complications related to catheter failure are also highlighted.

Keywords