EPJ Web of Conferences (Mar 2014)

Adaptation and security validation of an irradiator suitable for use of cesium-137 sources

  • Silva Rogério M.V.,
  • Belinato Walmir,
  • Santos William S.,
  • Junot Danilo O.,
  • Souza Luiza F.,
  • Souza Divanízia N.

DOI
https://doi.org/10.1051/epjconf/20146610016
Journal volume & issue
Vol. 66
p. 10016

Abstract

Read online

With the aim of offering a new application for 137Cs sources that were previously used in brachytherapy, we developed an irradiator for TL and OSL dosimeters by adapting a shielded storage container; this irradiator was named SOPHIA. It was assembled from a cubic iron casing, filled with lead, with external dimensions of 25 × 25 × 25 cm3. It has three different drawers with two sample holders suitable for samples or dosimeters with dimensions of up to 2 cm. Dose distribution of the surface of sample holder was obtained with EBT-3 radiochromic film and detectors thermoluminescent CaSO4: Dy. Regarding the security conditions of irradiation, the whole routine procedure for placement and removal of samples was simulated using Monte Carlo code. The equipment presented proper security conditions and the doses received by users during procedures for placement and extraction of the samples are in accordance with international standards.